Class III — Hamilton Medical AG

Recalling firmHamilton Medical AG
Registered facilityView this company's 3 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass III
StatusCompleted
ReasonWhen starting a humidifier, intended for respiratory gas conditioning during invasive and noninvasive mechanical ventilation, with an empty chamber, and water is not filled into the chamber, the low water level alarm will be delayed and medical staff are unaware that there is no water in the chamber.
ProductHAMILTON-H900 Humidifier, Models: 950001, 950004
Quantity203
DistributionUS: WI, FL, CA, NV, PA, TX, PA
Initiated2021-01-05
Classified2022-07-14
Recall numberZ-1385-2022
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)