Class III — Siemens Medical Solutions USA, Inc

Recalling firmSIEMENS MEDICAL SOLUTIONS USA, INC.
Registered facilityView this company's 9 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass III
StatusTerminated
ReasonThe Sensis Signal Input Box provides hooks as a milled item on the bottom of the box support the possibility of mounting the Signal Input Box on an accessory rail (e.g., at the patient table), In the event the mounting position deviates from the default position, the following are possible: Patient touching the Signal Input Box may result in a touch leakage current to the patient; Spillage of liquid entering the Signal Input Box
ProductAXIOM Sensis or Sensis Vibe Combo systems as follows: System Material # AXIOM Sensis, Combo 64 6623974; AXIOM Sensis, Hemo LOW 6634633; AXIOM Sensis, Combo 32 6634641; AXIOM Sensis, EP 129 6634658; Sensis 10764561; Sensis Vibe Combo 11007642; Sensis recording system is intended to be used as a diagnostic and administrative tool supporting hemodynamic catheterizations and/or electrophysiology studies, for cardiac as well as interventional Radiology.
Quantity399 units U.S.
DistributionUS Nationwide Distribution Foreign: Albania Algeria Angola Argentina Armenia Aruba Australia Austria Azerbaijan Bahrain Bangladesh Belgium Bolivia Bosnia and Herzegovina Brunei Darussalam Bulgaria Cambodia Cameroon Canada Chile China Colombia Costa Rica Croatia Cuba Cyprus Czech Republic Denmark Ecuador Egypt El Salvador Estonia Ethiopia Finland France Georgia…
Initiated2022-07-12
Classified2022-08-18
Terminated2024-11-07
Recall numberZ-1558-2022
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class III recall — Siemens Medical Solutions USA, Inc (Z-1558-2022).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/90632/

Not open yet

Alerts for SIEMENS MEDICAL SOLUTIONS USA, INC.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.