RegWardenFDA facility compliance records
Recall

Class II — Siemens Healthcare Diagnostics, Inc.

Recalling firmSIEMENS HEALTHCARE DIAGNOSTICS INC.
Registered facilityView this company's 12 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusOngoing
ReasonNegative bias with patient samples. The mean patient sample bias range from -26% to -33%,potential risk of delay in appropriate treatment and management decisions.
ProductDimension Vista LOCI CA 15-3 Flex reagent cartridge. In Vitro Diagnostic for the quantitative measurement of CA 15-3. Siemens Material Number (SMN)/Ref Catalog Number : 10464344/K6456
Quantity129 units US; 159 units OUS
DistributionWorldwide distribution - US Nationwide and the countries of Canada, Austria, France, Germany, Italy, Switzerland.
Initiated2022-07-18
Classified2022-08-29
Recall numberZ-1648-2022
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Siemens Healthcare Diagnostics, Inc. (Z-1648-2022).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/90669/

Not open yet

Alerts for SIEMENS HEALTHCARE DIAGNOSTICS INC.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.