RegWardenFDA facility compliance records
Recall

Class II — Zap Surgical Systems

Recalling firmZap Surgical Systems Inc.
Registered facilityZAP SURGICAL SYSTEMS INC. (the firm's only registered facility)
ClassificationClass II
StatusOngoing
ReasonSoftware issue identified in cases of initial patient setup with large (greater-than-or-equal-to 1.5 degrees) rotational deviations, between digitally reconstructed radiographs, and pre-delivery X-ray images. In these cases the transitional alignment algorithm may incorrectly calculate the new treatment table offset values, leading to an incorrect position for the subsequent treatment isocenter.
ProductZAP-X Radiosurgery System
Quantity8
DistributionWorldwide distribution - US Nationwide distribution in the states of AZ, MD, FL, CO and the countries of DEU, CHE, JPN.
Initiated2022-07-05
Classified2022-09-14
Recall numberZ-1742-2022
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Zap Surgical Systems (Z-1742-2022).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/90716/

Not open yet

Alerts for Zap Surgical Systems Inc.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.