Class II — Alcor Scientific, Inc.

Recalling firmALCOR Scientific LLC
Registered facilityALCOR SCIENTIFIC LLC (the firm's only registered facility)
ClassificationClass II
StatusOngoing
ReasonPrimary Pump Tubing Spare Part is incorrect. If the tubing is installed onto an ALCOR iSED ESR Analyzer, there is a potential of a biohazard leak from the iSEDs Probe Tube
ProductiSED Primary Pump Tubing Spare Part Kit- (IVD) tubing is used to move sample and wash solution from the aspiration needle, through the reading chamber, and to the waste on the iSED Automated ESR Analyzer PN: 112-13-002
Quantity38 units
DistributionUS Nationwide Distribution - FL, MA, MD, PA, SD, WI Foreign: Croatia Honduras Korea
Initiated2022-07-20
Classified2022-09-15
Recall numberZ-1751-2022
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)