RegWardenFDA facility compliance records
Recall

Class I — Fresenius Kabi USA, LLC

Recalling firmFresenius Kabi AG
Registered facilityView this company's 10 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass I
StatusTerminated
ReasonThe display screen may become frozen and unresponsive to user input, triggering an alarm that is evidenced by a buzzer sound and flashing red LEDs as a secondary notification. However, as the screen goes blank it does not indicate an alarm condition or respond. This issue leads to temporary cessation of flow of a medication or fluid, requiring immediate intervention by the clinician.
ProductIvenix Infusion System (IIS), LVP Software; LVP-SW-0004 version 5.2.0
Quantity4 instances
DistributionDomestic distribution to NJ and WI.
Initiated2022-09-08
Classified2022-10-06
Terminated2024-09-10
Recall numberZ-0002-2023
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class I recall — Fresenius Kabi USA, LLC (Z-0002-2023).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/90845/