RegWardenFDA facility compliance records
Recall

Class II — Belmont Instrument Corporation

Recalling firmBELMONT INSTRUMENT LLC.
Registered facilityBELMONT MEDICAL TECHNOLOGIES (the firm's only registered facility)
ClassificationClass II
StatusOngoing
ReasonPremature pump failure may lead to the device displaying a HALT 4 error, from which the device cannot recover even after device reboot. The issue may potentially lead to delayed treatment and a failure to deliver the prescribed thermoregulation therapy. To date, the firm has received one report of death potentially related to this issue.
ProductCritiCool, thermoregulation device
Quantity158 (US); 803 (OUS)
DistributionDomestic distribution nationwide. Foreign distribution worldwide.
Initiated2021-07-20
Classified2022-10-31
Recall numberZ-0153-2023
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Belmont Instrument Corporation (Z-0153-2023).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/90939/