RegWardenFDA facility compliance records
Recall

Class II — iRhythm Technologies, Inc.

Recalling firmiRhythm Technologies, Inc.
Registered facilityView this company's 2 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusOngoing
ReasonAmbulatory ECG monitoring system labeling update: 1) Clinically actionable arrhythmia notifications delayed until patient registration is complete; 2) Patches limited to 100 symptomatic and 500 asymptomatic wireless transmissions; once the maximum transmission limit is reached for either type, any further transmissions for that type will cease. Risk of delayed reporting to health providers.
ProductZio AT Clinical Reference Manual, Part: ALB0031, and Zio AT Important Information Pamphlet, Part: ALB0034. For Zio Ambulatory ECG monitoring system
Quantity1,443
DistributionUS: MD, NY, FL, TN, WA, AL, MA, AZ, CA, KS, HI, PA, NJ, TX, MN, VA, AR, OR, NC, CO, KY, MS, LA, MI, WI, WV, IN, ID, SC, CT, SD, IA, NV, OK, OH, IL, VT, NH, NE, NM, MO, ME, DC, RI, MT, DE, GA, ND
Initiated2022-09-28
Classified2022-11-04
Recall numberZ-0183-2023
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — iRhythm Technologies, Inc. (Z-0183-2023).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/90954/