Class II — Steris Corporation
| Recalling firm | STERIS Corporation |
|---|---|
| Registered facility | Key Surgical, LLC (the firm's only registered facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | Two electronic diode components (D1 & D4) placed in the reverse orientation on the controller printed circuit board assembly may result in premature wear of the power regulator, leading to the regulators failure which could cause a full power loss. If the regulator failure were to occur during a patient procedure, a procedural delay may result |
| Product | GI4000 Electrosurgical Unit - intended to deliver electrosurgical outputs to perform cutting and/or coagulation and coagulation with or without delivery of argon gas, in flexible endoscopic applications. MODEL #(s): G1110001 |
| Quantity | 5 units |
| Distribution | US Nationwide distribution in the states of Arizona, Illinois, Tennessee, Virginia. |
| Initiated | 2022-10-17 |
| Classified | 2022-11-16 |
| Terminated | 2024-05-28 |
| Recall number | Z-0241-2023 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Steris Corporation (Z-0241-2023).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/91020/
Alerts for STERIS Corporation
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?