RegWardenFDA facility compliance records
Recall

Class II — CooperSurgical, Inc.

Recalling firmCOOPERSURGICAL, INC.
Registered facilityView this company's 6 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusOngoing
ReasonFirm has received 11 complaints involving a loose or detached metal clips, including 3 cases where the metal clip was ingested by the patient, and 3 cases where the metal clip was removed from the patient's mouth. In one case, the patient suffered a laceration to the throat.
ProductNEO-fit Neonatal Endotracheal Tube Grip, Part Number 42-2540
Quantity220,500 devices
DistributionWorldwide distribution: US (Nationwide) including Puerto Rico; and countries (foreign) of: Australia, Bahrain, Belgium, Bulgaria, Canada, Colombia, Cyprus, Ecuador, Germany, Greece, Hong Kong, Hungary, India, Ireland, Italy, Malaysia, Malta, Netherlands, New Zealand, Saudi Arabia, Singapore, Turkey, Philippines, Qatar, United Arab Emirates, and United Kingdom.
Initiated2022-10-20
Classified2022-11-10
Recall numberZ-0223-2023
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — CooperSurgical, Inc. (Z-0223-2023).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/91047/