RegWardenFDA facility compliance records
Recall

Class II — Maquet Cardiovascular, LLC

Recalling firmMaquet Cardiovascular LLC
Registered facilityView this company's 2 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusOngoing
ReasonInk on the Vasoshield syringe Maquet logo may chip resulting in an unreasonable risk of harm to the patient. Potential harms may include foreign body reaction, coronary embolic event, and peripheral embolic event.
ProductHemopro 2 with Vasoshield, Model VH-4001
Quantity9817 devices
DistributionUS Nationwide. Japan, Germany.
Initiated2022-12-22
Classified2023-01-19
Recall numberZ-0995-2023
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Maquet Cardiovascular, LLC (Z-0995-2023).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/91363/