Class II — Randox Laboratories
| Recalling firm | RANDOX LABORATORIES, LTD. |
|---|---|
| Registered facility | View this company's 4 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Ongoing |
| Reason | The Antibody (R2 reagent) is showing a positive bias compared to previous batches on patient samples. Quality Controls do not show the same bias and will be within range |
| Product | CRP Immunoturbidimetric reagent. Intended for the quantitative in vitro determination of CRP concentration in serum Catalogue number: CP3826 |
| Quantity | 2 kits |
| Distribution | PA, OH, WV |
| Initiated | 2023-02-01 |
| Classified | 2023-03-24 |
| Recall number | Z-1286-2023 |
| Initiation | Voluntary: Firm initiated |