RegWardenFDA facility compliance records
Recall

Class II — Terumo Cardiovascular Systems Corporation

Recalling firmTERUMO CARDIOVASCULAR SYSTEMS CORP.
Registered facilityTERUMO CARDIOVASCULAR SYSTEMS CORPORATION (the firm's only registered facility)
ClassificationClass II
StatusOngoing
ReasonSome Capiox NX, RX and FX Oxygenators may contain an arterial thermistor component that is improperly bonded to the oxygenator port. This may result in a leak at the point where the arterial thermistor is improperly bonded to the oxygenator port.
ProductCAPIOX NX19 Hollow Fiber Oxygenator, Catalog Number 3CX*NX19RE
Quantity84 units
DistributionUS Nationwide. Belgium, Canada, Costa Rica, Thailand.
Initiated2023-03-01
Classified2023-04-05
Recall numberZ-1346-2023
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Terumo Cardiovascular Systems Corporation (Z-1346-2023).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/91775/

Not open yet

Alerts for TERUMO CARDIOVASCULAR SYSTEMS CORP.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.