Class II — T.A.G. MEDICAL PRODUCTS CORPORATION, LTD.
| Recalling firm | F & L MEDICAL PRODUCTS LLC |
|---|---|
| Registered facility | F & L MEDICAL PRODUCTS LLC (the firm's only registered facility) |
| Classification | Class II |
| Status | Ongoing |
| Reason | Drill guides might bend when axial forces are applied by surgeon during procedure and result in potential metal shavings due to friction with the drill bit.Risk to patients includes potential of injury. |
| Product | BETTA LINK SR REUSABLE FISHMOUTH GUIDE-Intended for use in soft tissue to bone fixation in the repair of the natural ligament or tendon disruption or to assist in reconstruction surgeries. Part Number: 110045151 |
| Quantity | N/A |
| Distribution | Worldwide distribution - US Nationwide and the countries of Australia, Japan, Netherlands. |
| Initiated | 2023-05-04 |
| Classified | 2023-07-03 |
| Recall number | Z-2063-2023 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — T.A.G. MEDICAL PRODUCTS CORPORATION, LTD. (Z-2063-2023).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/92410/
Alerts for F & L MEDICAL PRODUCTS LLC
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.