Class II — Siemens Healthcare Diagnostics, Inc.

Recalling firmSIEMENS HEALTHCARE DIAGNOSTICS INC.
Registered facilityView this company's 12 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusOngoing
ReasonA firmware error may lead to an incorrect association of test result(s) to a sample ID and therefore to an incorrect patient result. The mis-association of test results, or a delay in testing, may occur when all the conditions listed below occur within a few milliseconds timeframe of each other: - The SRM module is releasing a sample tube (Tube A) just placed into the carrier - Another sample tube (Tube B) is erroneously not diverted into the SRM module buffer - There is divert gate malfunction and a narrow timing window Only in this specific scenario, Tube A may be released by the module as Tube B due to a miscommunication between the module firmware and the Automation software without a unique error message being generated. The Automation System loses the traceability of Tube A. It manages both Tube A (incorrectly identified as Tube B) and the real Tube B according to the pending test orders (not yet performed) on Tube B.
ProductAptio Automation Storage and Retrieval Module (SRM)
Quantity83 units
DistributionWorldwide distribution - US Nationwide distribution in the states of CA, CO, FL, IL, KY, LA, MA, MD, MI, MS, MT, NC, NH, NY, OH, PA, SC, TN, TX, VA, WA, & WV. The countries of Australia, Brazil, Canada, China, France, Germany, Greece, Italy, Portugal, Saudi Arabia, Spain, Sweden, & United Kingdom.
Initiated2023-05-16
Classified2023-07-05
Recall numberZ-2091-2023
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Siemens Healthcare Diagnostics, Inc. (Z-2091-2023).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/92461/

Not open yet

Alerts for SIEMENS HEALTHCARE DIAGNOSTICS INC.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.