Class I — Datascope Corp.
| Recalling firm | Datascope Corp. |
|---|---|
| Registered facility | View this company's 3 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class I |
| Status | Ongoing |
| Reason | Users were reporting that the device was not charging as expected. It was discovered that users were unaware that the Cardiosave console was not completely inserted into the hospital cart. If the console is not fully inserted back into the cart the battery(ies) will not charge. |
| Product | Cardiosave Rescue Intra-Aortic Balloon Pumps (IABP), Part Numbers 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85 |
| Quantity | 9175 units |
| Distribution | Worldwide - US Nationwide distribution. |
| Initiated | 2023-07-31 |
| Classified | 2023-08-30 |
| Recall number | Z-2432-2023 |
| Initiation | Voluntary: Firm initiated |