RegWardenFDA facility compliance records
Recall

Class II — Ellex Medical Pty Ltd.

Recalling firmEllex Medical
Registered facilityEllex Medical Pty Ltd (the firm's only registered facility)
ClassificationClass II
StatusOngoing
ReasonWhen using a Laser-Indirect-Ophthalmoscope (LIO) as the delivery device, a scattered laser beam inside the laser cavity system may unintentionally exit through the Slit Lamp output port (SDS) and get emitted from the Objective Lens when the laser is fired.
ProductIntegre Pro Yellow LP561, Integre Pro Yellow LP6RY, Integre Pro Scan Green LP6G, Integre Pro Green LP5532 Spare, Optics Bench V2, L2G/LP6G
Quantity57 systems in total (6 units shipped to United States)
DistributionUS nationwide/ Worldwide Distribution
Initiated2023-03-16
Classified2023-06-27
Recall numberZ-1971-2023
InitiationFDA Mandated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Ellex Medical Pty Ltd. (Z-1971-2023).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/92532/