Class I — Baxter Healthcare Corporation
| Recalling firm | BAXTER HEALTHCARE CORPORATION |
|---|---|
| Registered facility | View this company's 32 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class I |
| Status | Ongoing |
| Reason | A medical device correction is being issued for SIGMA Spectrum Infusion System (V8 Platform) and Spectrum IQ Infusion System with Dose IQ Safety Software infusion pumps due to an increase in reported false upstream occlusion alarms following upgrades to software versions v8.01.01 and v9.02.01. |
| Product | SIGMA Spectrum Infusion Pump, Product Code 35700BAX2 |
| Quantity | 3306 units |
| Distribution | US Nationwide distribution. |
| Initiated | 2023-06-15 |
| Classified | 2023-07-13 |
| Recall number | Z-2103-2023 |
| Initiation | Voluntary: Firm initiated |