Class II — Beckman Coulter, Inc.
| Recalling firm | BECKMAN COULTER, INC. |
|---|---|
| Registered facility | View this company's 22 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Ongoing |
| Reason | It has been determined through an internal investigation that the alternate Access Thyroglobulin calibrator cards for use with the Thyg reagent packs distributed in April 2023 will cause a decrease of approximately 11% in patient results. |
| Product | Access Thyroglobulin Calibrators |
| Quantity | 5980 units |
| Distribution | worldwide except US, Canada, and China |
| Initiated | 2023-07-06 |
| Classified | 2023-10-11 |
| Recall number | Z-0100-2024 |
| Initiation | Voluntary: Firm initiated |