RegWardenFDA facility compliance records
Recall

Class II — Zimmer, Inc.

Recalling firmZIMMER, INC.
Registered facilityView this company's 10 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusOngoing
ReasonThread form issue of the locking holes in that the locking screws would not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock.
ProductZimmer Periarticular Locking Plate (ZPLP) Distal Lateral Femoral Plate 6 holes 159 mm Length-Indicated for temporary internal fixation and stabilization of osteotomies and fractures, including comminuted fractures, supracondylar fractures, intra-articular and extra-articular condylar fractures, fractures in osteopenic bone, nonunions, and malunions. Item Number: 00235710106
Quantity75 units
DistributionAR AZ MA ME MO NJ NM NY OR WA Foreign: Brazil China Dominican Republic Korea Netherlamds Taiwan
Initiated2023-06-27
Classified2023-07-28
Recall numberZ-2254-2023
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Zimmer, Inc. (Z-2254-2023).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/92681/