Class II — Siemens Medical Solutions USA, Inc.
| Recalling firm | SIEMENS MEDICAL SOLUTIONS USA, INC. |
|---|---|
| Registered facility | View this company's 9 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Ongoing |
| Reason | If a user-generated preset for an 18L6 transducer created on a 1.0 ultrasound system is used with a 2.0 ultrasound system, the 2.0 ultrasound system will display underestimated measurement results when using an 18L6 transducer and viewing in the Dual format visualization function, which may lead to misdiagnosis of a patient's condition or influence patient management decisions in a negative way. |
| Product | ACUSON Redwood 2.0 ultrasound systems, REF: 11503314 |
| Quantity | 796 |
| Distribution | Worldwide - US Nationwide distribution including in the states of IL, WI, CA, NM, TX, NC, MI, MD, NY, OH, IN, NE, PA, LA, SC, MO, OK, VA, ME, AZ, PR, IA, NH, NJ, AR, FL, SD and the countries of GR, ES, SG, TH, KR, IN, DE, TR, AT, IT, AU, JP, FR, CA, PT, SK, BR, SA, GB, NZ, RO, SE, AE, CN, HR, PL, CH, CR, IQ, NL, MX, ZA, BO, PH, MA, PA, EG, FJ, JO, DZ, TW, DK, CZ, CO, UG, CL, VN, CM, BG, QA, AZ, EC… |
| Initiated | 2023-07-13 |
| Classified | 2023-09-13 |
| Recall number | Z-2582-2023 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Siemens Medical Solutions USA, Inc. (Z-2582-2023).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/92750/
Alerts for SIEMENS MEDICAL SOLUTIONS USA, INC.
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.