Class II — AB Sciex, LLC
| Recalling firm | AB SCIEX |
|---|---|
| Registered facility | AB SCIEX LP (the firm's only registered facility) |
| Classification | Class II |
| Status | Ongoing |
| Reason | In rare occasions, the combination of a leak of a highly flammable solvent at the ion source probe, and an electrical discharge may result in a flame on top of the probe and possible ejection of the electrode assembly from the probe. There is a risk to the patient/user, customer property, or environment because there is a potential for a health risk and property damage should the flame be left unattended. The failure causes a delay in patient results. |
| Product | API3200MD Mass Spectrometer, Part Number 4466230 |
| Quantity | 3 |
| Distribution | US Nationwide. |
| Initiated | 2023-08-30 |
| Classified | 2023-10-05 |
| Recall number | Z-0037-2024 |
| Initiation | Voluntary: Firm initiated |