RegWardenFDA facility compliance records
Recall

Class II — LeMaitre Vascular, Inc.

Recalling firmLEMAITRE VASCULAR, INC.
Registered facilityView this company's 2 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusOngoing
ReasonBurrs (molding flash) on the cutter may tear the harvested vien
ProductThe Chevalier Valvulotome is a device used to disrupt the valves of veins which are to be used as grafts in the in situ or non-reversed bypass procedure. The Chevalier Valvulotome consists of a 70 cm stainless steel shaft with two tips: 1) the cutting tip is shaped like an inverted tulip with four teeth at its base. This tip is designed to cut the vein valve cusps. 2) the other tip is atraumatic and used to test the patency and the course of the vein before the valvulotomy. The Chevalier Valvulotome is available in three tip diameters: 2.5 mm (Catalog # 40025); 3.0 mm (Catalog # 40030); and…
Quantity225 units
DistributionInternational distribution to the countries of Belgium, France, Germany, and United Kingdom.
Initiated2023-07-13
Classified2023-08-28
Recall numberZ-2468-2023
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — LeMaitre Vascular, Inc. (Z-2468-2023).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/92825/

Not open yet

Alerts for LEMAITRE VASCULAR, INC.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.