Class II — PTW-FREIBURG
| Recalling firm | PTW Freiburg |
|---|---|
| Registered facility | PTW FREIBURG (the firm's only registered facility) |
| Classification | Class II |
| Status | Ongoing |
| Reason | Generated calibration file will not be correct in case of a relative calibration of an OCTAVIUS Detector 1000 or OCTAVIUS Detector 1600, if a Tiff file or a DICOM data set is used as reference matrix. Therefore, the measurement result will be incorrect by using such a calibration file. |
| Product | BeamAdjust software Version 2.3.3 Ref: S080032 The following devices are affected when a relative calibration is performed using ArrayCal (module from BeamAdjust S080032, UDI-DI: EPTWS0800320 ): " OCTAVIUS Detector 1600 SRS (T10056) (UDI-DI: EPTWT100560) " OCTAVIUS Detector 1600 MR (T10057) (UDI-DI: EPTWT100570) " OCTAVIUS Detector 1600 XDR (T10058) (UDI-DI: EPTWT100580) " OCTAVIUS Detector 1000 SRS (T10036) (UDI-DI: EPTWT100360) |
| Quantity | 42 US |
| Distribution | US Nationwide distribution. |
| Initiated | 2023-06-07 |
| Classified | 2023-09-11 |
| Recall number | Z-2577-2023 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — PTW-FREIBURG (Z-2577-2023).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/92879/