Class II — DKK Dai-Ichi Shomei Co., Ltd.

Recalling firmDKK DAI-ICHI SHOMEI CO, LTD
Registered facilityDKK DAI-ICHI SHOMEI CO, LTD (the firm's only registered facility)
ClassificationClass II
StatusOngoing
ReasonThe stopper block on the LFS bracket could be broken or missing. Without this stopper block, the flat screen attached to the LFS flat screen system can rotate beyond the original angle limit and it could cause excessive force to be given to the spring shaft of the LFS bracket. Due to the excessive force given to the spring shaft, the spring shaft could break and shear off.
ProductLFS Flat screen support arm systems without surgical lamp for models Aurora series, Aurora 2 series, Aurora 3 series, Aurora 4 series, Aurora Astro series, Stellar XL, and Stellar series, LFS Flat Screen Support Arm Model Numbers
Quantity1211 units
DistributionUS States: MI
Initiated2023-08-23
Classified2023-10-05
Recall numberZ-0072-2024
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — DKK Dai-Ichi Shomei Co., Ltd. (Z-0072-2024).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/92999/

Not open yet

Alerts for DKK DAI-ICHI SHOMEI CO, LTD

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.