RegWardenFDA facility compliance records
Recall

Class II — Encore Medical, LP

Recalling firmENCORE MEDICAL, L.P.
Registered facilityView this company's 3 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusOngoing
ReasonReverse shoulder prosthesis package that should contain 5mmx30 screws could contain incorrect 3.5mmx30 screws. Use of incorrect screws could result in surgical delay and inadequate fixation.
ProductREVERSE PROSTHESIS SHOULDER SYSTEM, SCREW, LOCKING BONE, RSP, 5mmX30mm long, REF: 506-03-130
Quantity70
DistributionUS: FL, IN, WI, IA, MN, KY, NY, AZ, RI, WA, OH, SC, VA, TX. OUS: MX
Initiated2023-08-22
Classified2023-10-12
Recall numberZ-0098-2024
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Encore Medical, LP (Z-0098-2024).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/93080/

Not open yet

Alerts for ENCORE MEDICAL, L.P.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.