RegWardenFDA facility compliance records
Recall

Class II — Abbott

Recalling firmABBOTT LABORATORIES
Registered facilityView this company's 64 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusOngoing
ReasonCertain NT2000iX RF generators were serviced using a tool that was subsequently found to be out of tolerance and therefore may not have provided accurate results for potential capacitance failure.
ProductAbbott NT2000iX Radiofrequency (RF) generator, REF RFG-NT-2000, lesioning generator
Quantity34 devices
DistributionUS, Taiwan, and Colombia
Initiated2023-09-28
Classified2023-11-02
Recall numberZ-0209-2024
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)