RegWardenFDA facility compliance records
Recall

Class I — B. Braun Medical, Inc.

Recalling firmB. BRAUN MEDICAL INC.
Registered facilityView this company's 5 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass I
StatusOngoing
ReasonIn affected products, the tubing outer diameter may vary in size or the length of tubing may be too short. Both issues may lead to false air-in-line alarms, which will stop the infusion and interrupt therapy. Interruptions or delays in medication administration may result in serious injury to patients which may be life-threatening (e.g. interruption of a vasopressor to a hemodynamically unstable patient) or may lead to death.
ProductInfusomat Administration Sets as listed below: 362031 Infusomat Space Pump Set with Caresite Injection Site; 362032 15 drops/mL, priming volume: 22mL, length: 119in, SafeLine; 362033 UNIV. 15 DR PUMP SET, 1.2 FIL, 2 CRSITE; 362034 Priming volume: 19mL, length: 120in, SafeLine; 362035 INFUSOMAT SPACE PUMP IV SET; 362043 Priming volume: 39mL, length: 120in, SafeLine; 362050 15 drops/mL, priming volume: 26mL, length: 124in, SafeLine; 362420 SPACE PUMP SET 15 DROP W/2 SAFEDAY; 362432 15 drops/mL, priming volume: 21mL, length: 118in, ULTRAS…
Quantity33,820,976 units
DistributionDistribution US nationwide, Canada and Germany.
Initiated2023-11-01
Classified2023-12-01
Recall numberZ-0355-2024
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class I recall — B. Braun Medical, Inc. (Z-0355-2024).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/93232/

Not open yet

Alerts for B. BRAUN MEDICAL INC.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.