RegWardenFDA facility compliance records
Recall

Class II — Elekta, Inc.

Recalling firmELEKTA AB
Registered facilityView this company's 4 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusOngoing
ReasonElekta has identified that if the microswitch within the middle arm is incorrectly setup, there is a potential for an uncontrolled extension of iViewGT / XVI detector arms. When the detector arm is fully deployed there is no risk of uncontrolled extension.
ProductELEKTA LIMITED Versa HD, REF XRT 2121, 2131; Medical Charged-Particle Radiation Therapy System Accelerator, Linear
Quantity451 units
Distributionworldwide
Initiated2023-10-25
Classified2023-12-01
Recall numberZ-0445-2024
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Elekta, Inc. (Z-0445-2024).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/93264/

Not open yet

Alerts for ELEKTA AB

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?