Class II — Sentinel CH SpA

Recalling firmSENTINEL CH. SpA
Registered facilitySENTINEL CH. SPA (the firm's only registered facility)
ClassificationClass II
StatusOngoing
ReasonA high recovery of Iron assay has been observed at some customer sites. Internal studies confirmed the positive shift of Quality Control and patient samples results, up to 15% for Iron Products (Alinity c Iron Reagent REF. 08P3920) to be used on Abbott Alinity Instrument and to 30% for Iron Products (IRON REF. 6K95-41 and IRON REF. 6K95- 30) to be used on Abbott Architect Instrument. When using the affected lots, there is a potential for delay in reporting patient results due to QC failure, also a potential for falsely elevated results.
ProductAlinity c Iron Reagent, Reference Number 08P3920
Quantity4400 kits
DistributionWorldwide - US Nationwide distribution.
Initiated2023-09-29
Classified2023-11-17
Recall numberZ-0326-2024
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Sentinel CH SpA (Z-0326-2024).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/93298/

Not open yet

Alerts for SENTINEL CH. SpA

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.