Class II — Breas Medical, Inc.
| Recalling firm | BREAS MEDICAL, INC. |
|---|---|
| Registered facility | BREAS MEDICAL, INC. (the firm's only registered facility) |
| Classification | Class II |
| Status | Ongoing |
| Reason | Produced with unintended open slits on the side of the mouthpiece body, may result in insufficient ventilation in spontaneous breathing patients using MPV-VCV and MPV-PCV modes of ventilation |
| Product | with Notch 22M/15F sold as part of Breas Medical MPV circuit 2 pc Kits: 006093, 006653, an accessory to ventilator device, Vivo 45LS. intended for partial insertion into a patient's mouth (in between the lips) to facilitate access to the respiratory system. |
| Quantity | 215 units |
| Distribution | Nationwide Foreign: Canada |
| Initiated | 2023-10-13 |
| Classified | 2023-11-27 |
| Recall number | Z-0405-2024 |
| Initiation | Voluntary: Firm initiated |