RegWardenFDA facility compliance records
Recall

Class II — Johnson & Johnson Surgical Vision, Inc.

Recalling firmJohnson & Johnson Surgical Vision, Inc.
Registered facilityView this company's 9 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusOngoing
ReasonDue to intraocular lens containing an angle out of specification (OOS). The cylinder axis marks do not meet product specifications for the allowable angle between Low Power Meridian (LPM) and toric cylinder axis marks.
ProductTECNIS Toric II OptiBlue IOL Models ZCW
Quantity36
DistributionInternational distribution to the country of Japan.
Initiated2023-12-12
Classified2024-01-16
Recall numberZ-0735-2024
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Johnson & Johnson Surgical Vision, Inc. (Z-0735-2024).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/93523/