Recall

Class II — Ambu Inc.

Recalling firmAmbu A/S
Registered facilityAMBU INNOVATION GMBH (the firm's only registered facility)
ClassificationClass II
StatusOngoing
ReasonInstructions for Use have been updated to add the following warning: Use only M4 screws with the length of 14 16 mm when mounting Ambu aView 2 Advance on a VESA interface. Using longer screw lengths will penetrate the lithium-ion battery and result in a fire hazard and battery leakage which can cause severe burns, smoke inhalation and skin irritation. Using shorter screw lengths could result in unsecure device fastening.
ProductAmbu aView 2 A Monitor, Catalog numbers 405011000 and 405011000US2
Quantity11845 distributed US
DistributionDomestic distribution nationwide.
Initiated2023-12-05
Classified2024-01-11
Recall numberZ-0714-2024
InitiationVoluntary: Firm initiated