RegWardenFDA facility compliance records
Recall

Class II — Maquet Cardiovascular, LLC

Recalling firmMaquet Cardiovascular LLC
Registered facilityView this company's 2 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusOngoing
ReasonRisk of polyimide particulate matter being present on the endoscopic harvesting vessel system. This issue may result in polyimide particulate matter being introduced into the arterial conduit for patients undergoing a coronary artery bypass graft (CABG) or peripheral bypass graft surgery.
ProductVasoview Hemopro Endoscopic Vessel Harvesting System, model VH-3000-W .
Quantity554 units (OUS)
DistributionUS: OUS: China, Brazil, Hong Kong.
Initiated2023-12-01
Classified2024-01-12
Recall numberZ-0723-2024
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Maquet Cardiovascular, LLC (Z-0723-2024).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/93615/

Not open yet

Alerts for Maquet Cardiovascular LLC

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.