RegWardenFDA facility compliance records
Recall

Class II — Cardinal Health 200, LLC

Recalling firmCardinal Health 200, LLC
Registered facilityView this company's 35 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusCompleted
ReasonThe above-listed lots of Cardinal Health Medi-Vac CRD" Suction Canister Liners 1L may experience loss of suction on low or low intermittent suction settings, due to a misalignment of the liner with the outer hard canister.
ProductCardinal Health MEDI-VAC CRD Suction Canister Liner, 1000 mL Suction Canister Liner Kit with 5 mm 1.8 m Tube, REF 65651-516
Quantity203244 units
DistributionUS, Canada, Australia and New Zealand
Initiated2023-12-21
Classified2024-01-18
Recall numberZ-0755-2024
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Cardinal Health 200, LLC (Z-0755-2024).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/93628/

Not open yet

Alerts for Cardinal Health 200, LLC

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.