Class II — Medtronic MiniMed, Inc.
| Recalling firm | Medtronic |
|---|---|
| Registered facility | View this company's 2 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Ongoing |
| Reason | Some sensors may have a glucose oxidase (GOX) layer thickness outside of specification (5.5 to 9.2 m). A non-conforming GOX layer thickness could impact sensor function (sensor glucose value) or loss of sensor function due to the sensor layer being too thin. The potential impact resulting from this issue ranges from user inconvenience to over/under delivery of insulin, possibly resulting in hyperglycemia or hypoglycemia, if the sensor glucose values are affected. The issue was identified by the manufacturing team during the execution of Filmetrics Measurement for a Guardian Sensor lot. |
| Product | Guardian 4 Glucose Sensor, REF: MMT-7040A, MMT-7040C1, MMT-7040C2, MMT-7040C3, MMT-7040D1, MMT-7040D2, MMT-7040MA, MMT-7040QC1, MMT-7040QC2, MMT-7040QC3 |
| Quantity | 114,758 |
| Distribution | Worldwide - US Nationwide distribution including in the states of PR, AK, AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY and the countries of Austria, Belgium, Canary Islands, Czech Republic, Denmark, Finland, France, Germany, Greece, Iceland, India, Ireland, Isra… |
| Initiated | 2023-11-15 |
| Classified | 2024-02-07 |
| Recall number | Z-1005-2024 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Medtronic MiniMed, Inc. (Z-1005-2024).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/93675/