RegWardenFDA facility compliance records
Recall

Class II — Accure Acne, Inc

Recalling firmAccure Acne, Inc.
Registered facilityACCURE ACNE, INC. (the firm's only registered facility)
ClassificationClass II
StatusOngoing
ReasonLaser system's spot tracking system may become misaligned resulting in the possible risk of delivery of higher-than-expected laser energy to epidermal tissue, which may cause blistering.
ProductAccure Laser System, Model: PFMS00004
Quantity22
DistributionUS distribution to states of: CO, CA, CT, NY, NJ, MN, LA, OH, MA, TX
Initiated2023-12-06
Classified2024-01-20
Recall numberZ-0778-2024
InitiationVoluntary: Firm initiated