Class I — Baxter Healthcare Corporation
| Recalling firm | BAXTER HEALTHCARE CORPORATION |
|---|---|
| Registered facility | View this company's 32 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class I |
| Status | Ongoing |
| Reason | An error was identified in software versions 2.0.8 and 2.1.8 while using the "Use Some Overfill" feature which may lead to over-delivery of an ingredient. This issue can result in redundant ingredient delivery of the requested overfill volume. |
| Product | Baxter Exactamix Pro 2400, REF EXM24DY |
| Quantity | 67 devices |
| Distribution | US: AL, AZ, CA, FL, GA, IA, KY, MD, MI, MN, MO, NC, NE, NH, NJ, NV, NY, OR, SD, TX, WA, WI, |
| Initiated | 2023-12-22 |
| Classified | 2024-02-16 |
| Recall number | Z-1002-2024 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class I recall — Baxter Healthcare Corporation (Z-1002-2024).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/93700/