Class II — Randox Laboratories Ltd.
| Recalling firm | RANDOX LABORATORIES, LTD. |
|---|---|
| Registered facility | View this company's 4 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Ongoing |
| Reason | Randox Laboratories have released an update to the Carryover Avoidance Technical Bulletin to detail that Copper, should not be run directly after Total Protein, on the RX series instruments testing order. If copper assay is run after total protein it could lead to falsely elevated copper results. |
| Product | RX Series Copper (Cu) Assay Ref. Number CU2340 |
| Quantity | 3,053 kits (2 US, 3,051 OUS) |
| Distribution | US Nationwide distribution in the states of SC and OK. |
| Initiated | 2023-11-29 |
| Classified | 2024-02-06 |
| Recall number | Z-0992-2024 |
| Initiation | Voluntary: Firm initiated |