Class II — Fresenius Kabi USA, LLC
| Recalling firm | Fresenius Kabi AG |
|---|---|
| Registered facility | View this company's 10 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Completed |
| Reason | Retroactively reported correction from 2022: A software defect may cause an incorrect (Fail-Stop) Alarm when an administration set is loaded with an upstream occlusion present, resulting in a non-recoverable (Fail-Stop) Pump Problem alarm. This may lead to delay of therapy, and patient may experience from no to minimal to serious harm. This issue was resolved in all fielded product in software version 5.2.1, which was installed in September 2022. |
| Product | LVP software of the Ivenix Infusion System (IIS), LVP-SW-0004 |
| Quantity | 3 |
| Distribution | US Nationwide distribution in the states of WI and NJ. |
| Initiated | 2022-09-16 |
| Classified | 2024-03-11 |
| Recall number | Z-1282-2024 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Fresenius Kabi USA, LLC (Z-1282-2024).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/93958/
Alerts for Fresenius Kabi AG