RegWardenFDA facility compliance records
Recall

Class II — Think Surgical, Inc.

Recalling firmThink Surgical Inc.
Registered facilityThink Surgical Inc. (the firm's only registered facility)
ClassificationClass II
StatusOngoing
ReasonThere is a galling effect (cold-weld) caused by friction between the array fixation pins and the TMINI array guide holes due to insufficient clearance between the outer diameter of the fixation pins and the inner diameter of the TMINI array guide holes. As a result, the array guide has the possibility of binding to the array fixation pins during drilling.
ProductAssembly, Array Guide TMINI, REF: 109101, part of the TMINI Knee Instrument Set, used with the TMINI Miniature Robotic System
Quantity16
DistributionUS Nationwide distribution in the states of AL, OH, NV, TX, UT, CA, MO.
Initiated2024-02-12
Classified2024-03-20
Recall numberZ-1318-2024
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Think Surgical, Inc. (Z-1318-2024).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/94086/

Not open yet

Alerts for Think Surgical Inc.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.