Class II — Exactech, Inc.
| Recalling firm | EXACTECH, INC. |
|---|---|
| Registered facility | View this company's 2 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Ongoing |
| Reason | The packaging of these affected UHMWPE humeral liners and glenoids are nonconforming as they do not meet the established packaging specification. They were packaged in vacuum bags that did not contain an additional oxygen barrier layer consisting of Ethylene Vinyl Alcohol (EVOH). |
| Product | Exactech Equinoxe GLENOID,POSTERIOR AUGMENT, PEGGED, 16o, LEFT, FOR CEMENTED USE ONLY, Mates with: a) 38, 41, 44, 47 head, Small, Item Number 314-06-22, b) 41, 44, 47, 50 head, Medium, Item Number 314-06-23, c) 44, 47, 50, 53 head, Large, Item Number 314-06-24, d) 44, 47, 50, 53 head, Extra Large, Item Number 314-06-25; Shoulder Arthroplasty |
| Quantity | 180 units |
| Distribution | Nationwide and Australia, Austria, Belgium, Brazil, Chile, Canada , China, Colombia, France, Germany, Great Britain, Greece, Guatemala, India, Italy, Japan, Korea, Luxembourg, Malaysia, Netherlands, New Zealand, Portugal, Singapore, South Africa, Spain, Sweden, Switzerland, Thailand |
| Initiated | 2024-03-06 |
| Classified | 2024-03-29 |
| Recall number | Z-1410-2024 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Exactech, Inc. (Z-1410-2024).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/94092/
Alerts for EXACTECH, INC.
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.