Recall

Class II — Agfa N.V.

Recalling firmAgfa N.V.
Registered facilityView this company's 2 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusOngoing
ReasonSteel cable installed inside the mobile column which supports the weight of the telescopic arm with the tube head has potential risk -an interference could cause the safety system which blocks the arm if the cable is cut off does not work properly, may result in patient injury
ProductAgfa DX-D 100 system-Digital Radiography mobile X-ray System. A mobile X-Ray modality, consisting of X-Ray console, generator, tube, digital detector, and workstation. This modality is used to generate projection X-Ray images of human patients. Models: 5411/ 050, A5411/ 0300 and A5411/ 0400
Quantity541 units
DistributionWorldwide distribution - US Nationwide and the countries of Canada, Argentina, Australia, Austria, Belgium, Chile, China, Colombia, Czech Republic, Finland, France, Germany, Greece, Italy, Kuwait, Luxembourg, Malaysia, Mexico, New Zealand, Norway, Peru, Poland, Romania, Russian Federation, Saudi Arabia, Singapore, Spain, Sweden, Switzerland, United Kingdom.
Initiated2024-02-23
Classified2024-04-02
Recall numberZ-1462-2024
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Agfa N.V. (Z-1462-2024).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/94143/

Not open yet

Alerts for Agfa N.V.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.