Class I — Outset Medical, Inc.
| Recalling firm | Outset Medical, Inc. |
|---|---|
| Registered facility | View this company's 3 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class I |
| Status | Ongoing |
| Reason | Hemodialysis System, pre and post dialyzer peroxide-cured silicone tubing, with less than 336 hours usage, will be replaced because tubing may contain non-dioxin-like (NDL) polychlorinated biphenyl acids (PCBAs), that have not been flushed out over time with usage. Exposure to PCBAs could cause: skin conditions, liver damage, impaired reproduction, development of certain kinds of cancer in humans. |
| Product | Tablo Hemodialysis System, REF: PN-0003000, PN-0006000 and PN-0007001 |
| Quantity | 919 |
| Distribution | US:FL, CA, NM, SC, AZ, TN, NC, TX, IL, MD, IN, DE, AL, MI, AK, KY, LA, ID, OK, IA, VA, GA, MA, NE, KS, WI, PA, OH, WA, MT, RI, MN, MO, AR, NY, CO, WV, SD, OR, UT, WY, DC. OUS: AE |
| Initiated | 2024-03-06 |
| Classified | 2024-04-12 |
| Recall number | Z-1485-2024 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class I recall — Outset Medical, Inc. (Z-1485-2024).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/94224/