Recall

Class II — FUJIFILM Healthcare Americas Corporation

Recalling firmFUJIFILM Healthcare Americas Corporation
Registered facilityFUJIFILM Healthcare Americas Corporation (the firm's only registered facility)
ClassificationClass II
StatusOngoing
ReasonAC adapters of Noblus can heat up, and cause smoke in the worst case, when an excessive mechanical stress is applied to the socket for power cable (frequent insertion/removal of the power cable).
ProductNoblus AC Adapter used with the Noblus Ultrasound Imaging System Model : AHM250PS24
Quantity418 units
DistributionUS Nationwide distribution.
Initiated2024-04-11
Classified2024-05-22
Recall numberZ-1884-2024
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)