Class II — Cadwell Industries Inc

Recalling firmCADWELL INDUSTRIES, INC.
Registered facilityCADWELL INDUSTRIES, INC. (the firm's only registered facility)
ClassificationClass II
StatusOngoing
ReasonThere is the potential that an electroneurodiagnostic medical device insulator may be installed incorrectly which could result in arcing to patient connections resulting in potential unintended shock to the patient.
ProductCascade IOMAX Cortical Module, REF: 190296-200
Quantity54 units
DistributionWorldwide - US Nationwide distribution in the state of MD and the countries of Mexico, Canada, Malaysia, India, Taiwan, Saudi Arabia, European Union, Switzerland, South Africa, Iraq, United Kingdom.
Initiated2024-04-23
Classified2024-05-22
Recall numberZ-1882-2024
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)