RegWardenFDA facility compliance records
Recall

Class II — Covidien, LLC

Recalling firmCovidien llc
Registered facilityView this company's 22 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusOngoing
ReasonDue to saline conductivity issues, their is a potential that dilation catheters may provide inaccurate esophageal diameter measurements that may lead to various harms.
ProductEsoFLIP, 30mm BALLOON DILATION CATHETER, REF ES-330, Rx Only
QuantityN/A
DistributionWorldwide - US Nationwide distribution including in the states of AK, AL, AZ, CA, CO, CT, DC, FL, GA, IA, IL, KS, KY, LA, MA, MD, MI, MN, MO, NC, NE, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI and the countries of Argentina, Australia, Austria, Bahrain, Belgium, Canada, Denmark, Germany, Ireland, Italy, Kuwait, New Zealand, Norway, Portugal, Qatar, South Africa, Switzerland, United …
Initiated2024-04-25
Classified2024-06-07
Recall numberZ-2038-2024
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Covidien, LLC (Z-2038-2024).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/94541/

Not open yet

Alerts for Covidien llc

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.