Class II — Stryker Neurovascular

Recalling firmSTRYKER CORP.
Registered facilityView this company's 61 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusOngoing
ReasonStryker Neurovascular has observed an increased frequency of PTFE coating damage occurring on the Synchro Guidewires that may be caused by the practice of backloading the guidewire through the optional introducer accessory. This issue is limited to certain lots of the Synchro Guidewire that contain an older version of the introducer accessory. Due to variation in the manufacturing process of the supplier of the introducer accessory, certain lots of introducers have sharper than intended edges that can peel off the PTFE coating when physicians use a technique known as backloading. Users with impacted product in their inventory are cautioned not to use this backloading technique.
ProductThe Synchro SELECT Guidewires are a steerable guidewire family with a shapeable tip and are available in straight and pre-shaped versions. The outside diameters of the guidewires are 0.014in. The guidewires are compatible with existing microcatheters used in common procedures such as those used in endovascular diagnosis and therapy of neurovascular disease. The distal portion of the guidewire tip is radiopaque. For lubricity, the distal portion of the device is coated with a hydrophilic polymer and the proximal portion of the guidewire is coated with polytetrafluoroethylene (PTFE). The torq…
Quantity41,074
DistributionWorldwide - US Nationwide distribution including in the states of TX, NY, MD, CA, KY, WI, OR, NE, SC, MA, MO, CT, ID, PA, TN, FL, NC, AL, IA, OK, VA, IL, OH, NH, MN, ND, KS, NJ, WV, IN, GA, HI, AR, UT, LA, MI, WA, NV, CO, AZ, DC, VT, ME, DE, AK, WY, MT, NM, SD, MS, RI, GU and the countries of ARGENTINA, AUSTRALIA, AUSTRIA, BAHRAIN, BELGIUM, BRAZIL, BULGARIA, CANADA, CHINA, COLOMBIA, CROATIA, CZECH…
Initiated2024-04-18
Classified2024-05-31
Recall numberZ-1940-2024
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Stryker Neurovascular (Z-1940-2024).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/94543/

Not open yet

Alerts for STRYKER CORP.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.