RegWardenFDA facility compliance records
Recall

Class II — Echonous Inc

Recalling firmECHONOUS, INC.
Registered facilityECHONOUS, INC. (the firm's only registered facility)
ClassificationClass II
StatusOngoing
ReasonUltrasound system has a bug in affected iOS software, which after DICOM export, can cause inaccuracy in postprocessed measurements done in 3rd party software, such as PACS or other post-processing software, which may lead to inaccurate diagnoses or treatment.
ProductKosmos on iOS impacting iOS, Model Numbers: P007639, P007944, software is part of a general-purpose diagnostic ultrasound system.
Quantity272
DistributionUS: ID, TX, WI, KS, IL, GA, MI, CA, OH, FL, AK, UT, MD, NY, NV, CO, ND, WA, DC, LA, KY, MN, HI, NM, TN, NJ, ID, MI, VT, AZ, MO, AL, MA, NC OUS: Denmark, Switzerland, Austria, Italy, France, United Kingdom, Germany, Spain, Romania, Finland, Romania, Australia, India, Hong Kong, Thailand
Initiated2024-05-09
Classified2024-07-10
Recall numberZ-2325-2024
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Echonous Inc (Z-2325-2024).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/94709/

Not open yet

Alerts for ECHONOUS, INC.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.