RegWardenFDA facility compliance records
Recall

Class II — Angiodynamics, Inc.

Recalling firmANGIODYNAMICS, INC
Registered facilityView this company's 4 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusOngoing
ReasonA single batch of ISO-100 IsoLoc Gas-Release Endorectal Balloons is mislabeled. The affected ISO-100 balloon catheter has a nominal fill volume of 100cc/ml but the inner pouch identifies the product as an ISO-60 (fill volume of 60cc/ml).
ProductIsoLoc Prostate Immobilization Treatment Balloon Device Kit, Catalog/UPN: ISO-100
Quantity1,500 units (75 kits with 20 units per kit)
DistributionUS Nationwide distribution in the states of AZ, FL, IL, MO, MS, NJ, NM, NY, PA, TX, WI.
Initiated2024-05-21
Classified2024-07-03
Recall numberZ-2267-2024
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Angiodynamics, Inc. (Z-2267-2024).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/94773/